Trigger modes and appendant registers

Channel Trigger

(c) Spectrum GmbH

71

Channel pulsewidth trigger for short positive pulses

The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from lower to higher values (rising
edge) the pulsewidth counter is started. If the pulsewidth
counter reaches the programmed amount of samples, no
trigger will be detected.

If the signal does cross the triggerlevel again within the the
programmed pulsewidth time, a triggerevent will be detect-
ed.

Channel pulsewidth trigger for short negative pulses

The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from higher to lower values (falling
edge) the pulsewidth counter is started. If the pulsewidth
counter reaches the programmed amount of samples, no
trigger will be detected.
If the signal does cross the triggerlevel again within the the
programmed pulsewidth time, a triggerevent will be detect-
ed.

Register

Value

Direction

set to

Value

SPC_TRIGGERMODE

40000

r/w

TM_CHANNEL

20040

SPC_TRIGGERMODE0

40200

r/w

TM_CHXPOSG_SP

10002

SPC_HIGHLEVEL0

42000

r/w

Set it to the desired triggerlevel relatively to the channel’s input range.

board dependant

SPC_PULSEWIDTH

44000

r/w

Set to the desired pulsewidth in samples.

2 to 255

Register

Value

Direction

set to

Value

SPC_TRIGGERMODE

40000

r/w

TM_CHANNEL

20040

SPC_TRIGGERMODE0

40200

r/w

TM_CHXNEG_SP

10012

SPC_HIGHLEVEL0

42000

r/w

Set it to the desired triggerlevel relatively to the channel’s input range.

board dependant

SPC_PULSEWIDTH

44000

r/w

Set to the desired pulsewidth in samples.

2 to 255