7 environmental specifications, Environmental specifications – Symmetricom Time Server User Manual

Page 18

Advertising
background image

1-4

TimeVault™ User’s Manual

6000-100Ch1.fm Rev. D

Chapter 1: General Information

Environmental Specifications

1.7

Environmental Specifications

The environmental specifications of TimeVault are:

Operating Temperature

TimeVault Module:

0 to +50 °C (+32 to +122 °F)

Antenna:

–40 to +70 °C (–40 to +158° F)

Maximum Rate of Change:

8 °C per hour

Storage Temperature

TimeVault Module:

–50 to +85 °C (–40 to +185 °F)

Antenna:

–55 to +85 °C (–67 to +185 °F)

Maximum Rate of Change:

15 °C per hour

Operating Humidity

TimeVault Module:

0% up to 95%, non-condensing

Antenna:

100%, condensing

Storage Humidity

TimeVault Module:

0% up to 95%, non-condensing

Antenna:

100%, condensing

Operating Altitude

TimeVault Module:

Maximum 4 km

Storage Altitude

TimeVault Module:

Maximum 12 km

Shock & Vibration Requirements

In Shipping Container:

Per ISTA Procedure 2A

Bench Handling without Shipping
Container:

Per EN60068-2-31