28

930317 Rev. C

I X .

S E T U P ,

A D J U S T M E N T & U S E

Note on Torque Setting:

A torque setting is

the optimum tightening which should be made
on a particular fastener.A torque wrench
which measures in./lbs. Should be used when
tightening fasteners on Quickie wheelchairs.
Unless otherwise noted, use a torque setting of
120 in./lbs.When setting up this wheelchair.

A. BATTERY REMOVAL AND

INSTALLATION
To remove the batteries for transporting

or servicing:

1. Remove Batteries

a) Unplug the battery connectors located

on the controller plate for both

battery boxes.

b) Push battery cables under battery

tray lever (A).

c) Lift tray lever (A) up.

d) Fold tailgate (B) down.

e) Slide batteries under motors, away

from chair.

f) Lift up batteries by lifting up on

battery lifting straps (C ).

Batteries my weigh up to 55 lbs. Care

must be taken to avoid injury when lifting.

2. Installing Batteries

a) Lift tray lever (A) up, lower tailgate (B).

b) Slide batteries onto tailgate. Cables

should be positioned to the rear

of batteries.

c) Slide batteries all the way forward on

battery tray.

d) Lift up on tailgate.

e) Pull tray lever down.

f) Plug battery cables into connectors.

B

C

A