N specifications, Nspecifications – Konica Minolta Non-Contact 3D Digitizer Range7 User Manual

Page 23

Advertising
background image

21

n

Specifications

MODEL NAME

KONICA MINOLTA RANGE7

Measuring Method

Triangulation by light sectioning method

Light-Source

Semiconductor laser λ = 660 nm

Laser Classification

Class 2 (IEC60825-1 edition2)

Sensor

Name

CMOS Photo diode

Pixel

1.31 Mega Pixel (1280 x 1024)

Measuring Distance (mm)

450 to 800 (Distance from Light-Receiving Lens *1)

Light Receiving Lens (Replaceable)

TELE / WIDE

Measuring
Range
(mm)

MultiFocus

mode

Disabled

Lens

TELE

TELE

WIDE

WIDE

Distance

450

800

450

800

X-Y size

79 Ч 99

141 Ч 176

150 Ч 188

267 Ч 334

Z

54

97

109

194

Meas.pitch of X-Y

0.08

0.14

0.16

0.28

MultiFocus

mode

Enabled

Lens

TELE

TELE

WIDE

WIDE

Distance

462

781

475

766

X-Y size

81 Ч 102

138 Ч 172

159 Ч 199

256 Ч 320

Z

54

97

109

194

Meas.pitch of X-Y

0.08

0.13

0.08

0.25

Accuracy (Interglobuler distance)*2

±40 μm

Precision (Z, σ)*3

4 μm

Auto Focus

available

Auto Exposure

available

Scanning Time

about 2 sec. per scan

Preview Function

available (about 0.4 sec. per scan)

Environment Brightness

500 lx or less

File Format *4

INPUT/OUTPUT: .rgv, .rvm, .rmk (Original format), OUT only: .stl

Output Interface

USB 2.0 HighSpeed

Power Supply

AC adapter
100-240Vac (50-60 Hz), 1.4A

External Dimensions (mm)

295 (W) x 190 (H) x 200 (D)
* exclude the protruding portion of the lens and handle

Weight (kg)

about 6.7

Operating Condition

10 to 40ºC (relative humidity 65% or less, with no condensation)

Storage Condition

-10 to 50ºC (relative humidity 85% or less [at 35ºC] , with no
condensation)

Remarks (Controlled Method)

Controlled via dedicated PC software "RANGE VIEWER"

*1 Due to the difference of origins on the measurement reference surface and measurement data, the Z axis coordinate has an offset of about 120 mm for the measuring

distance.

*2 When the interglobuler distance for the ball bars (2 globes) specified in VDI/VDE2634-2 is measured under the following conditions defined by KONICA

MINOLTA OPTICS:
Our conditions: Temperature 20±1ºC / Using the TELE lens / Distance 450 mm / Warming up 20 mins. / Using the KONICA MINOLTA OPTICS processing

software / With Calibration / Object to be measured: Standard KONICA MINOLTA OPTICS (2 globes) instrument / Position of object to be
measured: KONICA MINOLTA OPTICS standard position (10 locations within the measurement space) / Does not cover the uncertainty over the
pricing of the standard instrument.

*3 Measuring conditions: Temperature 20±1ºC / Using the TELE lens / Distance 450 mm / Warming up 20 mins. / Using the KONICA MINOLTA OPTICS processing

software / Object to be measured: KONICA MINOLTA OPTICS reference plain chart / 1σ

*4 Using the KONICA MINOLTA OPTICS “RANGE VIEWER” processing software

• The RANGE7 includes the eT-Kernel/Compact from sSOL Co., Ltd.