Maintaining the Equipment

4-2

May 2002

Cleaning the Light Bar

If artifacts appear on your scanned images, clean the light bar.

IMPORTANT: Wear white cotton gloves for this procedure to prevent oils and dirt

from being absorbed into the microfiber cleaning cloth.

1. Fold a dry, untreated microfiber cleaning cloth so the edge of the cloth is

about the thickness of the light bar.

NOTE: You can use an e-wipe, manufactured by Photographic Solutions, Inc.,

instead of a microfiber cleaning cloth.

2. Gently move the edge of the microfiber cloth or e-wipe down the length of the

light bar one time.

3. Calibrate the scanner.

4. If an artifact remains after cleaning the light bar, check the light bar with a

magnifying glass for damage, such as a scratch or a chip. If there is damage
to the light bar, contact your Kodak representative for service.

Replacement Procedures

Replacing the Lamp

White cotton gloves are required for this procedure.

One spare lamp is provided with the scanner.

IMPORTANT: Use only the GE ELC Lamp with the HR Film Scanner (see

Appendix B: Ordering Accessories and Supplies).

It is not necessary to power down the scanner before changing the lamp.

1. Open the lamp-and-filter-access door.

P192_0016HC

lamp-and-filter-
access door

lift door open from here