Using the Service and Assembly Module (SAM)

A-8

May 2002

Calibrating the Scanner

IMPORTANT: The CAL (70MM-CAL for HR 500 Plus Film Scanner and 4 x 5 for

HR Universal Film Scanner) film holder must be in position in the
scanner before you continue. See Placing the Empty Film Holder in
the Film Holder Guide on page 2-6
.

1. If the values on the Calibrate/Image Processing tab do not match the values

shown on the sample on page A-7, change them. The sample is a
representation of the default values.

2. Select Shift Image Data Bits Up by Four.

3. Click Calibrate All.

Using the Capture Tab

From the Operations window, click the Capture tab when you need to perform a
scan.