Procedural error, Control solution that has not been shaken enough, Expired or contaminated control solution – Lifescan OneTouch Basic Basic Blood Glucose Monitoring System User Manual

Page 31: Improper coding of the meter, Test strip deterioration, Expired or contaminated test strip

Advertising
background image

Control solution test results are
automatically stored with the word
contrl. (If you use a logbook, indicate
control solution tests by writing “C”
after the result.)

Control solution results

If test results fall outside the range
printed on the test strip vial, repeat the
test. Results that fall outside the
expected range may indicate:

Procedural error

Control solution that has not been
shaken enough

Expired or contaminated control
solution

Improper coding of the meter

Dirt on test strip holder or on test
area beneath the test strip holder

Test strip deterioration

Expired or contaminated test strip

Do not use after expiration date. us