Comparing the results:

When the lab tests your blood, only
the serum or plasma portion is used.
Your OneTouch

®

Basic

®

Meter uses

only whole blood.

Comparing serum or plasma to whole
blood is like comparing a temperature
of 98.6°F with the temperature of
37°C—in order to see if the values are
equal, you have to adjust one of the
results.

Whole blood results are
approximately 12% lower than serum
or plasma results.

To compare lab results with meter
results, divide the lab result by 1.12.

Example:

Assume that your lab result (plasma or
serum) is 224 mg/dL.

224 divided by 1.12 = 200 (the lab
value is now approximately equal to a
whole blood result).

Compare this figure to your meter
result. The two should be within 20%
of each other.

You may still have a variation from the
result because blood glucose levels can
change significantly over short periods,
especially if you have recently eaten,
exercised, taken medication, or
experienced stress.