Load mps #1 rear view mps #2 rear view, Setup & connections – Lakeshore Learning Materials 622 User Manual

Page 24

Advertising
background image

Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Setup & Connections

2-8

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

+

-

+

-

+

-

OVP Im Vm m Vp +Vs -ls lp +ls -S +S

CAL AND ID

V

l

MODE

INT

V

l

EXT

RI

FLT

ON

WARNING:

NO USER SERVICEABLE PARTS INSIDE. REFER
SERVICING TO TRAINED SERVICE PERSONNEL.

208

220

240

100

120

-10% +5% 50-60 Hz

FIXED INTERNALLY

LINE

N

L

!

-OUT

+OUT

10K

10K

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

+

-

+

-

+

-

OVP Im Vm m Vp +Vs -ls lp +ls -S +S

CAL AND ID

V

l

MODE

INT

V

l

EXT

RI

FLT

ON

WARNING:

NO USER SERVICEABLE PARTS INSIDE. REFER
SERVICING TO TRAINED SERVICE PERSONNEL.

208

220

240

100

120

-10% +5% 50-60 Hz

FIXED INTERNALLY

LINE

N

L

!

-OUT

+OUT

10K

10K

+

-

LOAD

MPS #1 Rear View

MPS #2 Rear View

Figure 2-4. Typical Multiple MPS Connections for Two MPS Units

2.6.1

Multiple MPS Remote Inhibit Mode

In multiple MPS configuration, if the MPS RI indicators activate from an external contact closure, the MPS

units enter Multiple MPS Remote Inhibit Mode. They activate output overvoltage protection circuits and turn off

output circuits. If a charged load is present, OVP circuits discharge it. The front panel Fault annunciators turn

ON, and the internal audio indicators beep about once per second. If a MPS detects a fault, it initiates Multiple

MPS Remote Inhibit Mode by closing the FLT contacts. No delay is allowed (CAL AND ID switch 7 OFF, down

position) and multiple MPS operation (CAL AND ID switch 4 ON, up position) must be selected. After the load

discharges, turn OFF all units to reset.