Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Options & Accessories

6-5

6.2.5.2

Field Probe Setting and Calibration

Press Function Menu and then Next Menu to
display Function Menu 2. Press the FIELD Function
key to display the Field screen to the right.

FIELD PROBE SETTING:
UNITS: T or kG

Field Units. The Data Entry up or down arrows scroll

between the units: T (tesla) or kG (kilogauss). The default is kG.

CALIBRATE PROBE:

NOTE: The entries below change probe calibration. Do not change unless probe is actually being calibrated.

mV/kG: +XXX.XXX

Field Probe Sensitivity. Each Hall Probe comes with a mV/kG sensitivity. Use any numeric entry mode to

change the value from 0 to 999.999.

ZERO RDG: <OFF> or <ON>

Field Probe Zero Reading. Connect the Hall Probe and place it in zero magnetic field. Press a Data Entry up

or down arrow key to zero the probe.

6.2.6

Level Sensor Calibration

To enter or measure level sensor parameters,
select CALIBRATE SENSOR (line 8) in the LIQUID
HELIUM LEVEL screen. Prior to this, select a length
of inches or centimeters (not percent). Units of
length do not display on the calibration screen if
percent is selected.

LENGTH: 24.0 in or LENGTH: 61.0 cm

Level Sensor Length. Use any numeric entry mode to change the value. Allows values from 0.0 to 999.9.

MIN

: 0.00 (ohms)

Level Sensor minimum resistance in ohms. If minimum resistance is known, enter it directly with the Data

Entry keys. Otherwise, measure it as shown below under SELF CALIBRATION: CALIBRATE MIN

.

/inch: 10.41 or

/cm : 4.10

Level Sensor resistance per unit length. If the value is known, enter it directly with the Data Entry keys.

Otherwise, it can be measured as shown below under SELF CALIBRATION: CALIBRATE

/inch(cm).

CALIBRATE MIN

¥

¥

¥

¥

Measures the Level Sensor minimum resistance. Connect the level sensor immersed in LHe to the active level

sensor length. Move the cursor to CALIBRATE MIN

and press a Data Entry up/down arrow key. That line of

the display changes to CALIBRATING MIN

. After measurement, the display updates the minimum

resistance in line 3 and the display returns to that shown above.

CALIBRATE

/inch ¥

¥

¥

¥

or CALIBRATE

/cm ¥

¥

¥

¥

Measures the Level Sensor resistance per unit length. The resistance per unit length is a function of

temperature. The effective value can be determined from the value measured at a more convenient

temperature. Figure 6-1 and Table 6-2 show the resistance values as a function of temperature of a NbTi

filament. The procedure described below measures the resistance per unit length at room temperature.

Take the level sensor out of the liquid and allow its temperature to stabilize. Move the cursor to CALIBRATE

/inch and press a Data Entry up/down arrow key. That line of the display changes to CALIBRATING

/inch(cm)…. After measurement, the display updates the

/inch (

/cm) in line 4 and the display returns to

that shown above.

NOTE: If the resistance is too large, the value is not updated. In that case, use a lower temperature or an

ohmmeter to measure the resistance.

F I E L D P R O B E S E T T I N G :

¤ U N I T S

C A L I B R A T E P R O B E

m V / k G

Z E R O R D G

O U T P U T S :

*

+1.999A

: k G ¥

: + 0 . 8 0 0

: ¥ T O C A L I B R A T E

L V L : + 1 4 . 4 i n

F L D : + 0 . 5 0 1 k G

+ 0 . 0 0 1 V

L E V E L S E N S O R C O N S T A N T S :

¤ L E N G T H

M I N ¦

¦ / i n c h

S E L F C A L I B R A T I O N :

C A L I B R A T E M I N ¦

C A L I B R A T E ¦ / i n c h

O U T P U T S :

*

+0.000A

: 2 4 . 0 i n

: 0 . 0 0

: 1 0 . 4 1

L V L : + 1 4 . 4 i n

F L D : + 0 . 5 0 1 k G

+0.000V

¥

¥