Important Control Solution Test Information

• Use only OneTouch

®

Ultra

Control Solution.

• Check the expiration date on the control solution bottle. Record

the discard date (date opened plus three months) on the bottle
label. Do not use after expiration or discard date, whichever
comes first.

• Control solution, meter, and test strips should be at room

temperature between 20 and 25°C before testing.

• Shake the bottle, discard the first drop of control solution, and

wipe off the dispenser tip to ensure a good sample and an
accurate result.

• Store the control solution tightly closed at temperatures below

30°C. Do not refrigerate.

CAUTION: The control solution range printed on the test strip vial
is for OneTouch

®

Ultra

Control Solution only. It is not a

recommended range for your blood glucose level.