COMPARING METER AND
LABORATORY RESULTS

OneTouch

®

UltraSmart

Meter test results and laboratory test

results are both expressed in plasma-equivalent units. However,
your meter result may differ somewhat from your laboratory result
due to normal variation. Meter results can be affected by factors
and conditions that do not affect laboratory results in the same
way. Your OneTouch

®

UltraSmart

Meter glucose value should

agree with a laboratory measurement within ±20% most of the
time under normal conditions. A result within that range is
considered accurate when testing with the OneTouch

®

UltraSmart

Meter. However, results can differ by more than ±20% in some
situations. See the OneTouch

®

Ultra

Test Strip package insert for

typical accuracy and precision data and for important information
on “Limitations of Procedure.”

To make an accurate comparison between meter and laboratory
results, follow these guidelines:

Before you go to the lab:

• Perform a control solution test to make sure the meter is

working properly.

• It is best to fast for at least eight hours before doing

comparison tests.

• Take your meter with you to the lab.