MESSAGE

WHAT IT MEANS

The meter has
detected that the
temperature is
outside the system
operating range. Do
not perform a test
until the meter and
test strips reach a
temperature within
the operating range
of 6–44°C.

There may be a
problem with the
meter or the meter
may need to be reset.

ACTION

Perform the test after
the meter and test
strips have reached
a temperature within
the operating range.

To reset the meter, you
must turn the meter
off and enter set-up
mode (see pages 54-
57). Reset the
language, time, and
date. You must also
reset the meter code
to match the code on
the test strip vial.
If you continue to get
an Error 1 message,
do not use the meter.
Call your local
LifeScan Customer
Service office. See
back cover for contact
information.

WARNING

See Owner's Booklet

Test not possible

Meter is outside of

operating

temperature range

WARNING

See Owner's Booklet

Error 1

Call customer service

DCO_06158108A_OUS_OB_GB 5/6/03 1:25 PM Page 48