The status register, The a/d software trigger register, 4 the status register – Omega 1002 User Manual

Page 31: 5 the a/d software trigger register

Advertising
background image

3.2.2.4 The Status Register

Address 10h is used is the status register. Reading from this address will

return the data from the status register. The format of status register is:

Bit7-6 Bit5 Bit4 Bit3 Bit2 Bit1 Bit0

Gain

Control

8245

Timer 1

8245

Timer 0

8245

Timer 2

Reserved

Analog

input type

A/D Busy

Bit 7-6: Current A/D gain control.
Bit 5 : Output of 8254 timer 1.
Bit 4 : Output of 8254 timer 0.
Bit 3 : Output of 8254 timer 2.
Bit 2 : Reserved. Used for hardware testing.
Bit 1 : Analog input type, ‘1’ indicated that analog input type is single-

ended and ‘0’ indicated analog input is differential.

Bit 0 : A/D busy signal. ‘0’ indicates busy, A/D is under conversion. ‘1’

indicates not busy, A/D is complete conversion and is idle now.

3.2.2.5 The A/D Software Trigger Register


Writing to port 1Ch will generate an A/D trigger signal.

Note: Since the user can trigger at a rate greater than the speed of A/D converter

(125K), a delay time may be required between successive software triggers








A/D
Busy

Software
trigger

8

µs

Delay time

Conversion Time

Figure 3-1. Software triggered delay time.