Omega FC-21 User Manual

Page 32

Advertising
background image

28

FC-21 Flow Computer

Sub-menus

Display

Notes

6.4.11
SETUP AUX INPUT

NOTE:
When Density (Dens)
is selected, The menu
prompts will be very
similar to the Tem-
perature prompts.
The menus will
prompt the user for
density values and
density units.

SETUP AUX INPUT

AUX INPUT TYPE
None Dens

Temp

AUX SIGNAL TYPE
Voltage Current RTD

INPUT SIGNAL RANGE
0-10V 0-5V 1-5V

INPUT SIGNAL RANGE
4-20mA 0-20mA

AUX LOW SCALE
###### F

AUX FULL SCALE
###### F

AUX DEFAULT
###### F

AUX LOW ALARM
###### F

AUX HIGH ALARM
###### F

DENS EXTRACT METHOD
Therm_Coef API_2540

Advance To

SET FLUID PROPERTIES

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

Press ENTER to begin setup of the Auxiliary
Input.

Select Temperature to set the Auxiliary Input
for Temperature inputs.

Choose Temperature Signal Type.
Advance to "Aux Default", if RTD selected.

If Voltage selected,
Choose desired Voltage Range. Skip if RTD.

If Current selected,
Choose desired Current Range. Skip if RTD.

Enter the low temperature scale
corresponding to the low temperature signal.
Skip if RTD.

Enter the high temperature scale
corresponding to the high temperature signal.
Skip if RTD.

Enter the Default Temperature.
The unit will use this value if the temperature
input fails.

Enter the Low setpoint for the Temperature
Alarm.

Enter the High setpoint for the Temperature
Alarm.

Choose the Density Extraction method to be
used.

ENTER

Advertising