Omega FC-21 User Manual

Page 53

Advertising
background image

49

FC-21 Flow Computer

Sub-menus

Display

Notes

Calibrate ch1 0V
Vin=TB1-2 GND=TB1-4

Calibrate ch1 0V
0 CALIBRATING ——

Calibrate ch1 0V
*** DONE ***

Calibrate ch1 0V
Iin=TB1-2 GND=TB1-4

Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4

Calibrate ch1 10V
0 CALIBRATING ——

Calibrate ch1 10V
*** DONE ***

Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4

Advance to

Calibrate ch2 0V

8.2.10
Calibrate CH1 0V
Submenu Group

To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 0V and press Enter.

This message is displayed during calibration.

This message is displayed when the 0V
calibration is finished.

The display will automatically return to the
Calibrate CH1 0V submenu. Press the Down
arrow key to advance to the CH1 10V
calibration.

To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 10V and press Enter.

This message is displayed during calibration.

This message is displayed when the 10V
calibration is finished.

The display will automatically return to the
Calibrate CH1 10V submenu. Press the
Down arrow key to advance to the CH2 0V
calibration.

ENTER

STOP

ENTER

STOP

8.2.11
Calibrate CH1 10V
Submenu Group