FC-21 Flow Computer

Sub-menus

Display

Notes

Pulse input test

Pulse input test
Trigger level 2.5V

Pulse input test
count speed 3kHz

Pulse input test
F1: 0 F2: 0

Pulse input test

Analog out test

Analog out test
*0 4 10 15 20 mA

Analog out test

Excitation out test

Excitation out test
*5v 12v 24v

Excitation out test

Press Enter key to test the pulse input.

Use the Up/Down arrow keys to select the
appropriate trigger level.

Use the Up/Down arrow keys to select the
appropriate frequency range.

To check Pulse input accuracy: Use TB1-4 as
reference ground, input a frequency on TB1-2.
The display should show frequency being
input. Use a frequency counter to verify input.

Press Menu key to return to Pulse input test
top-level menu.

Press Enter to test the analog output.

To simulate analog output: Connect an
ammeter to (+) TB1-15, (-) TB1-16. Press the
key under the desired setting to move the
asterisk (*). The unit should output the
selected current.

Press Menu key to return to Analog out test
top-level menu.

Press Enter to test the excitation output.

To test the excitation output: Connect a
voltmeter to (+) TB1-1, (-) TB1-4. Press the
key under the desired setting to move the
asterisk (*). The unit should output the
selected voltage.

Press Menu key to return to Excitation out test
top-level menu.

8.2.18
Pulse input test
Submenu Group

ENTER

ENTER

ENTER

MENU

ENTER

MENU

8.2.19
Analog out test
Submenu Group

8.2.20
Excitation out test
Submenu Group

ENTER

MENU

STOP

START

STOP

START

2.5V
10mV

100mV

40Hz
3KHz

20kHz