3

English

Please ensure that you read the entire guide before using the Humidifier.

Indications for use

The Humidifier is indicated for humidification of the air delivered from a
ResMed compatible CPAP, bilevel or non-invasive ventilation device. This
device is for use only as recommended by a physician.

Contraindications

This device is contraindicated for use with patients whose upper
(supraglottic) airway has been bypassed. Please also refer to the relevant
system manual for the contraindications associated with Positive Airway
Pressure (PAP) therapy.

Humidifier parts

Refer to Illustration A on page 1.
Your heated Humidifier is made up of two main parts:
(1) docking station and (2) water chamber. Your docking station has the
following items (3) air outlet; (4) cover; (5) cover seal; (6) control dial; (7)
connector seal; (8) heater plate; (9) latch; (10) indicator light. Your water
chamber includes the (11) spill guard.

Setting up your humidifier for use

Refer to illustrations B & C on page 1.
Remove the connector plug from your flow generator (a) and attach it to
the back of the docking station (b).
1. Align your Humidifier with your flow generator and push together

until they click into place.

2. Connect the power cord to the socket on the rear of your flow

generator.

3. Plug the other end into a power outlet. Place the combined unit on

a firm, level surface near your bed at a level lower than your sleeping
position. Ensure the surface is not sensitive to heat and/or moisture, or
use a protective mat.

4. Connect one end of the air tubing firmly onto the air outlet.
5. Connect the assembled mask system to the free end of the air tubing.

WARNING

• Make sure your flow generator is disconnected from the power

outlet before connecting to docking station.

• The Humidifier docking station should only be connected or

disconnected when the water chamber is empty.

• Do not transport the flow generator with the Humidifier attached.
• The H4i is contraindicated for invasive use. An external humidifier

approved for invasive use is recommended according to EN ISO

8185.

Filling the water chamber

Refer to illustration D on page 1.
1. Open the cover.
2. Remove the water chamber.
3. Fill with fresh, clean water up to or below the maximum water level

mark.
Note: The water chamber must be removed from the H4i before it is filled
or water may enter the flow generator. The spill guard must also be in
place.

4. Return the water chamber to the docking station.

Note: Ensure that no water is collected between the heater plate and the
water chamber.

5. Close the cover ensuring that it clicks into place.

WARNING

Do not overfill the water chamber as water may enter the flow

generator and air tubing.

CAUTION

Do not use any additives (eg, scented oils or perfumes). They may

reduce the humidification output of the Humidifier and/or cause

deterioration of the water chamber materials.

Warming your humidifier with the VPAP or S8

Series

(Not available on all ResMed PAP devices)
1. Select warmup by pressing

on the control panel.

2. The WARMING screen will appear to indicate that the Warm-Up feature

is activated.

3. While your Humidifier is in Warm-Up mode you can:

• Start treatment at any time by pressing
• Access menus by pressing

• Stop the Warm-Up feature at any time by pressing

.

Warming your humidifier with the Stellar

Your Humidifier will be automatically detected when the Stellar is turned
on. The treatment screen will give you the option to start warming up
the Humidifier prior to starting treatment. If the Humidifier is heating, the
related symbol is displayed at the top of the LCD screen.
Note: The Humidifier can only be used when the device is connected to
mains power supply.

CAUTION

The Humidifier cannot be used for treatment pressures exceeding

35 cm H

2

O.

Starting your treatment

1. Turn the humidity control dial to three ( ). You can adjust the

control dial up or down at any time to find the setting that is most
comfortable for you. The

setting switches heated humidification

off and delivers minimum humidification while a setting of six ( )
delivers maximum humidification.

2. Start treatment as described in your device user guide.

Note: The indicator light will brighten to indicate when the device is
heating and dim when it has reached its set temperature.

3. Lie down and arrange the tubing so that it moves freely when you

turn in your sleep.

WARNING

Do not leave an excessive length of air tubing around the top of your

bed. It could twist around your head or neck while you are sleeping.

Stopping your treatment

To stop your treatment, refer to your device user guide.

Detaching your humidifier

Refer to illustration E page 1.
1. Unplug the power cord from the power outlet.

WARNING

Take care when handling your Humidifier. The water and water

chamber base may be hot. Allow 10 minutes for the heater plate and

any excess water to cool.

2. Remove the water chamber from the docking station.
3. Press the large silver button located midway underneath the docking

station and pull the Humidifier and flow generator apart.

Cleaning and maintenance

You should regularly carry out cleaning and maintenance of the water
chamber, cover and cover seal as described in this section.

WARNING

Beware of electric shock. Do not immerse your flow generator,

docking station or power cord in water. Always unplug the device

before cleaning and be sure that it is dry before plugging it back in.

Cleaning the water chamber

Refer to illustration F on page 2.
Daily
Wash the assembled water chamber, spill guard, tub plate and tub plate
seal in warm water, using a mild detergent. Rinse thoroughly in clean
water and allow them to dry away from direct sunlight.

Disassembling the cleanable water chamber

Refer to illustration G on page 2.
1. Remove the spill guard by pushing up from inside the air inlet and

pulling it away from the water chamber.