Activeforever Aeroneb Go User Manual

Page 3

Advertising
background image

IMPORTANT SAFEGUARDS

When you use electrical products, especially around children,
always follow basic safety precautions.

READ ALL INSTRUCTIONS BEFORE YOU USE YOUR NEBULIZER

WARNING:

• The AC/DC Adapter does not charge batteries.

• If you are using batteries and the green light is flashing, your

batteries are low. You may finish your treatment and then
change the three "AA" batteries immediately.

• Do not share your Aeroneb

®

Go Nebulizer with others. It is

intended for use by a single user. If more than one person
uses it, there is a risk of spreading infectious illness.

• Always unplug this product immediately when not in use.

• Adult supervision is required when this product is used by

individuals who require special assistance, such as children.

• Use this product only for physician-prescribed solutions for

inhalation as described in this manual.

• Do not use components or attachments that are not

recommended by

evo

Medical Solutions. Refer to the

Replacement Parts and Accessories section in this manual.

• Never operate this product if any of the parts are not working

properly or have been damaged.

• Do not use the Aeroneb

®

Go Nebulizer in the presence of a

flammable anesthetic mixture with air, oxygen or nitrous oxide.

• Do not use the Aeroneb

®

Go Nebulizer with anesthetic

breathing systems or lung ventilator breathing systems.

17

2

PERFORMANCE

• Aerosol flow rate: Greater than 0.25 mL per minute
• Residual Volume: <0.1 mL for a 3 mL dose
• Aerosol output rate: 0.2 mg per minute (salbutamol)
• Average particle size: 3.6 microns mass median

aerodynamic diameter

• Medication Cup capacity: Maximum of 6 mL
• Average pneumatic resistance @60 L/min: 2.2 cm/L/sec
• Rebreathed volume with mouthpiece: 46 mL
• The temperature of the medication may rise to 10°C (18°F)

above ambient temperature during normal use.

Performance may vary depending on the type of drug you use
and may not apply to drugs not approved for general purpose
nebulizers such as suspensions or drugs with high viscosity.
For additional information, contact

evo

Medical Solutions or

your drug supplier.

Performance may also be affected by improper cleaning.
Carefully clean the Medication Cup after every use. See
Cleaning Instructions on page 11.

Advertising