IMPORTANT CAUTION

As with any electro-mechanical device, this product may
become unusable due to an electrical outage, battery
depletion, or mechanical failure. We recommend that you have
spare batteries and have a backup device available to you.

CLASSIFICATIONS AND MARKINGS

The symbols that appear on the Aeroneb

®

Go Nebulizer and its

accessories are described in the table below:

TROUBLESHOOTING GUIDE

If you have difficulty operating your Aeroneb

®

Go, check the

following points:

If above remedies do not work, call your local supplier.

PROBLEM

CHECK

REMEDY

No visible aerosol
when using batteries

Check that the
batteries are inserted
correctly.

Using the markings under
each battery location,
place each battery in the
Battery Powered
Controller housing.
Match (+) and (-).

Check to see if the
low battery power
light is flashing.

Replace the batteries or
use the AC Powered
Controller.

Check that the Battery
Powered Controller
cable is connected
properly to the
nebulizer and to the
Control Module.

Connect the cable
properly to the nebulizer
and the Battery Powered
Controller.
(Refer to page 7).

No visible aerosol
when using AC
Powered Controller.

Check that the AC
Powered Controller is
plugged correctly
into a working AC
outlet.

Insert the plug into a
working outlet and verify
that the green light on
the controller is lit (Refer
to page 9).

Check that the AC
Powered Controller
cable is connected
properly to the
Aeroneb

®

Go

Nebulizer Unit.

Connect the cable
properly to the nebulizer
and the AC Powered
Controller.
(Refer to page 9).

No visible aerosol
when power source
is functioning

Check that there is
medication in the
Medication Cup.

Fill the Medication
Cup with prescribed
medication.

Inspect Medication
Cup to see if it is dirty.

Clean the Aeroneb

®

Go.

Refer to page 11.