Low Priority Alerts (continued)

Condition/Action/Explanation

Battery error has occurred. Check the connection of your

battery and ensure that it is properly attached and latched

on concentrator. If battery error recurs with same battery,

stop using the battery and switch to a new battery or

remove battery and operate concentrator using external

power supply.

Concentrator is producing oxygen at a slightly low level

(<82%) for a period of 10 minutes. If condition persists,

contact your equipment provider.

Battery has exceeded its charging temperature and charging

has stopped. The battery will not charge while this alert

is present but will begin to charge when the battery

temperature returns to the normal operating range.

If battery charging is desired sooner, remove the battery

from the concentrator and allow it to cool in an open area for

approximately 10-15 minutes. Then, re-insert the battery into

the Inogen One® G3. If the problem still persists, contact your

equipment provider.

Concentrator is producing oxygen but cannot report

battery status. Replace battery. If condition persists,

contact your equipment provider.

The concentrator requires servicing at the earliest

convenience. The concentrator is operating to specification

and may continue to be used. Contact your equipment

provider to arrange for service.

The concentrator’s oxygen sensor has malfunctioned.

You may continue to use the concentrator. If the condition

persists, contact your equipment provider.

Message Display & Text

Check Battery

Oxygen Low

See Manual

Remove Battery

to Cool

Comm Error

See Manual

Service Needed

O2 Sensor Fail

See Manual