Medium priority alerts – Activeforever Inogen One G3 Portable Oxygen Concentrator User Manual

Page 30

Advertising
background image

26

Condition/Action/Explanation

Concentrator has not detected a breath for 60 seconds.

Check that cannula is connected to concentrator, there are

no kinks in tubing and cannula is positioned properly in

your nose.

Oxygen output concentration has been below 50% for

10 minutes. If condition persists, switch to your backup

oxygen source and contact your equipment provider to

arrange for service.

A breath has been recognized, but proper oxygen delivery

has not been detected.

Concentrator is experiencing an error but is capable of

continued operation. Remove and re-insert battery and/

or check external power supply connections. If condition

persists, contact your equipment provider.

Battery has exceeded temperature limit while concentrator

is running on battery power. If possible, move concentrator

to a cooler location or power unit with an external power

supply and remove battery. If condition persists, contact

your equipment provider.

Message Display & Text

No Breath Detect

Check Cannula

Oxygen Error

Service Needed

O2 Delivery Error

System Error

See Manual

Battery HOT

Warning

Medium Priority Alerts

The following medium priority alert messages are accompanied by a triple beep,

repeated every 25 seconds, and a flashing red light.

www.ActiveForever.com