Inogen One® G3 System Specifications Chapter 8

English

ELECTrOMAGNETIC COMPATIBILITY
This CE Marked equipment has been tested and found to comply with the EMC limits for

the Medical Device Directive 93/42/EEC [EN 55011 Class B and EN 60601-1-2]. These limits

are designed to provide reasonable protection against harmful interference in a typical

medical installation.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The Inogen One® G3 Oxygen Concentrator is intended for use in the electromagnetic

environment specified below. The customer or the user of the Inogen One® G3 Oxygen

Concentrator should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions

CISPR 11

Group 1

The Inogen One® G3 Oxygen

Concentrator uses RF energy only for

its internal function. Therefore its RF

emissions are very low and not likely

to cause any interference in nearby

equipment.

RF emissions

CISPR 11

Class B

The Inogen One® G3 Oxygen

Concentrator is suitable for use in all

establishments, including domestic

establishments and those directly

connected to the public low-voltage

power supply network that supplies

buildings used for domestic purposes.

Harmonic Emissions

IEC 61000-3-2

Class A

Voltage fluctuations /

flicker emissions

IEC 61000-3-3

Complies