Cleaning, Care and Maintenance Chapter 6

English

4. Install a replacement fuse, Inogen RP#125 (BUSS MDA-12) and reassemble the tip.

Ensure the retainer ring is properly seated and tightened.

Inogen One® G3 Column Change Procedure

NOTE

Column change instructions are only to be used when maintenance is

required and are not intended for practice purpose.

1. Turn off the Inogen One® G3 concentrator by pressing the power button to shut

down the device.

2. Remove the Inogen One® G3 concentrator from the carry bag.

3. Remove the battery from the Inogen One® G3 concentrator.

4. Turn Inogen One® G3 over so that is standing in the

upside down position where the display panel is

facing down and the bottom of the Inogen One® G3

is facing up.

5. There are two columns (metal tubes) on the

Inogen One® G3 concentrator and one is located on

each side of the device.

Fuse

Cigarette Adapter Plug

retainer Tip

retainer

CAUTION For continued protection against risk of fire, only use fuse specified.