Filter Cleaning and replacement

Both particle filters must be cleaned weekly to ensure the ease

of air flow. Remove filters from the front of the device. Clean the

particle filters with a mild liquid detergent (such as Dawn

TM

) and

water; rinse in water and dry before reuse.

NOTE

It may be necessary to clean the particle filters more often in dusty

environments.

To purchase additional particle filters contact your equipment provider or Inogen.

Output Filter

The output filter is intended to protect the user from inhalation of small particles in the

product gas flow. The Inogen One® G3 includes an output filter conveniently located

behind the removable cannula nozzle fitting. Inogen requires that this filter be replaced

between patients.

The output filter may be replaced by the equipment provider or by the owner using the

Output Filter Replacement Kit (RP-107).
The Inogen One G3 Concentrator must be cleaned and disinfected as per the above

instructions for each new patient. No special maintenance needs to be carried out by the

patient. Your equipment provider performs maintenance operations to assure continued

reliable service from your Inogen One G3. The manufacturer’s instructions for the

preventative maintenance of the devices are defined in the service manual. All work must

be performed by trained technicians certified by the manufacturer.

DC Input Cable Fuse replacement

The Cigarette Lighter DC power plug contains a fuse. If the DC input cable is being used

with a known good power source and the unit is not receiving power, the fuse may need

to be replaced.

To replace the fuse, follow these instructions and refer to the photograph below.

1. Remove the tip by unscrewing the retainer. Use a tool if necessary.

2. Remove the retainer, tip and fuse.

3. The spring should remain inside the Cigarette Lighter Adapter housing. If the spring is

removed, make sure to replace the spring first before inserting the replacement fuse.