Inogen One® G3 Column Change Procedure (continued)

11. Insert column (metal tube) into the Inogen One® G3

concentrator. Do not leave the column exposed;

it should be inserted into the Inogen One G3 as

soon as the dust caps have been removed.

12. Push the column (metal tube) into the device

such that the column (metal tube) is fully seated

into the Inogen One® G3 concentrator. The spring

loaded latch button should fully return to the

closed position.

13. Repeat steps 10-12 to install the other column

(metal tube).

14. Connect the AC power supply cord to the Inogen

One® G3 and plug the power supply AC cord into an

electrical outlet. Do Not Power on the Inogen

One® G3 concentrator.

15. Press and hold the light button for 10 seconds and

screen will display “sieve reset” message. Release

button once message is displayed on screen.

16. Press the bell button once and screen will display “sieve reset successful”.

17. Press the power button to turn on the Inogen One® G3, and use normally.

wArNING Do not use any columns other than those specified in this user manual.

The use of non-specified columns may create a safety hazard and/or

impair equipment performance and will void your warranty.

!