VEGA VEGABAR 65 Profibus PA User Manual
Page 56
Nominal range
Overload, max. pressure Overload, min. pressure
-0.7 … 0.7 psig
200 psig
-15 psig
-1.5 … 1.5 psig
290 psig
-15 psig
-3 … 3 psig
430 psig
-15 psig
-7 … 7 psig
500 psig
-15 psig
Absolute pressure
0 … 15 psi
500 psi
0 psi
0 … 35 psi
700 psi
0 psi
0 … 70 psi
900 psi
0 psi
0 … 150 psi
1300 psi
0 psi
0 … 350 psi
1900 psi
0 psi
Reference conditions and actuating variables (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
-
Temperature
18 … 30 °C (64 … 86 °F)
-
Relative humidity
45 … 75 %
-
Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Determination of characteristics
limit point adjustment according to DIN 16086
Characteristics
Linear
Reference installation position
upright, diaphragm points downward
Influence of the installation position
<
5 mbar/0.5 kPa (0.07 psig)
Deviation determined according to the limit point method according to IEC 60770
11
)
Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue
current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation
-
Turn down 1 : 1 up to 5 : 1
<
0.075 %
-
Turn down > 5 : 1
<
0.015 % x TD
Deviation with absolute pressure measuring range 0.1 bar
-
Turn down 1 : 1 up to 5 : 1
<
0.25 %
-
Turn down > 5 : 1
<
0.05 % x TD
Influence of the product or ambient temperature
Applies to instruments with digital signal output (HART, Profibus PA, Foundation Fieldbus) as well
as to instruments with analogue current output 4 … 20 mA. Specifications refer to the set span.
Turn down (TD) = nominal measuring range/set span.
Thermal change zero signal and output span, reference temperature 20 °C (68 °F):
11)
Incl. non-linearity, hysteresis and non-repeatability.
56
VEGABAR
65 • Profibus PA
10 Supplement
28253
-EN
-090522