VEGA VEGABAR 65 Profibus PA User Manual

Page 56

Advertising
background image

Nominal range

Overload, max. pressure Overload, min. pressure

-0.7 … 0.7 psig

200 psig

-15 psig

-1.5 … 1.5 psig

290 psig

-15 psig

-3 … 3 psig

430 psig

-15 psig

-7 … 7 psig

500 psig

-15 psig

Absolute pressure

0 … 15 psi

500 psi

0 psi

0 … 35 psi

700 psi

0 psi

0 … 70 psi

900 psi

0 psi

0 … 150 psi

1300 psi

0 psi

0 … 350 psi

1900 psi

0 psi

Reference conditions and actuating variables (similar to DIN EN 60770-1)

Reference conditions according to DIN EN 61298-1

-

Temperature

18 … 30 °C (64 … 86 °F)

-

Relative humidity

45 … 75 %

-

Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Determination of characteristics

limit point adjustment according to DIN 16086

Characteristics

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

<

5 mbar/0.5 kPa (0.07 psig)

Deviation determined according to the limit point method according to IEC 60770

11

)

Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue
current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.

Deviation

-

Turn down 1 : 1 up to 5 : 1

<

0.075 %

-

Turn down > 5 : 1

<

0.015 % x TD

Deviation with absolute pressure measuring range 0.1 bar

-

Turn down 1 : 1 up to 5 : 1

<

0.25 %

-

Turn down > 5 : 1

<

0.05 % x TD

Influence of the product or ambient temperature
Applies to instruments with digital signal output (HART, Profibus PA, Foundation Fieldbus) as well
as to instruments with analogue current output 4 … 20 mA. Specifications refer to the set span.
Turn down (TD) = nominal measuring range/set span.

Thermal change zero signal and output span, reference temperature 20 °C (68 °F):

11)

Incl. non-linearity, hysteresis and non-repeatability.

56

VEGABAR

65 • Profibus PA

10 Supplement

28253

-EN

-090522