VEGA VEGASWING 83 fail safe and signal conditioning instrument VEGATOR 526 Ex User Manual

Page 10

Advertising
background image

10

VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe

Product description

Relay output (fail safe)

Number

1 switching relay (closing contact, floating)

Mode

A/B-switch
A - max. detection or overfill protection
B - min. detection or dry run protection

Contact

1 spdt each

Contact material

AgNi and Au plated

Turn-on voltage

min. 12 V DC
max. 250 V AC, 48 V DC

Switching current

min. 10 mA DC
max. 2 A AC, 1 A DC

Breaking capacity

max. 125 VA AC, 60 W DC

Fail safe relay (not fail safe)

Number

1 fail safe relay (spdt, floating)

Contact material

AgCd and au plated

Turn-on voltage

min. 10 mV
max. 250 V AC, 48 V DC

Switching current

min. 10

µ

A

max. 2 A AC, 1 A DC

Breaking capacity

max. 125 VA AC, 54 W DC

Indicating elements

LED in front plate
- green on

operating voltage on

- yellow

switch point control

- red

fault signal

Set-up key (green)

ready for set-up

Approvals

Max. values

U

O

= 18,9 V; I

K

= 95 mA; P = 647 mW

Characteristics

linear

EEx ia IIC

EEx ia IIB

Max. permissible outer inductance (mH)

0,5

1

1,5

2

0,5

1

1,5

2

Max. permissible outer capacitance (nF)

52

40

27

15

200

190

177

160

EEx ib IIC

EEx ib IIB

Max. permissible outer inductance

4 mH

15 mH

Max. permissible outer capacitance

240 nF

1,1

µ

F

The intrinsically safe circuits are reliably galvanically isolated from all not-intrinsically safe
circuits up to a peak value of 375␣ V of the nominal voltage.

Advertising