3 error rectification – VEGA VEGAFLEX 52P (Profibus PA) User Manual
Page 48
48
VEGAFLEX 51P and 52P
5.3 Error rectification
Error
Error rectification
E 013
Sensor cannot find
- Message is displayed during the warm-up
level echo
phase.
- If the message remains, the DK-value
of the medium could be too low.
Check the useful level and noise level.
See 4.2 Adjustment with adjustment module
MINICOM / Useful and noise level (6).
If the message still remains, carry out a
new adjustment.
E 017
Adjustment span too small
Carry out a new adjustment.
Make sure that the difference between min.
and max. adjustment is at least 10 mm.
E 036
Software update incorrect
Return the instrument for repair.
- Sensor requires programming (service).
- Failure message remains also during a
current programming.
E 040
Hardware failure /
Check all connection cables.
electronics defect
Contact our service department.
Diagnostics