Capacitive electrodes EL 4 … 20 mA - Compact

22652-EN-020809

Setup

4 Setup

4.1 General adjustment

The numbers in brackets relate to the illustra-
tion under “3 Electrical connection“. In the
setup procedure, the electrode must be ad-
justed to the medium which will later be meas-
ured.
In certain cases, a dry adjustment is also
possible.

The electrode can be adjusted in three differ-
ent ways:
• with the integrated oscillator
• with the adjustment software VEGA Visual

Operating (VVO from version 2.30)

• with a HART

®

handheld

CAP E 32 Ex

Adjustment

- directly on the oscillator.

CAP E 32 H Ex

Adjustment

- directly on the oscillator
- via PC with the adjustment

program VVO

1)

- via HART

®

handheld

Oscillator

The capacitive electrodes EL can be adjusted
directly on the oscillators CAP E32 Ex and
CAP E32 H Ex.
All basic sensor functions can be executed via
the two keys and the rotary switch.

PC with adjustment program VVO

1)

The capacitive electrodes, in conjunction with
oscillator CAP E 32 H Ex, can be very conven-
iently operated with the adjustment program
VVO (VEGA Visual Operating). Further func-
tions are available beside the sensor basic
functions.

For this, you require an interface adapter
VEGACONNECT 2 which you can connect
directly to any position on the signal cable or
directly to the sensor.

System requirements:
- IBM compatible PC with a serial port. We

recommend a PC with Pentium processor
with a clock frequency of
100 MHz.

- Memory: 16 MB
- Software requirements: Windows 95

HART

®

handheld

The capacitive electrodes EL with oscillator
CAP E32 H Ex are HART

®

protocol capable

and can be adjusted with a HART

®

handheld.

All relevant sensor functions can be executed
with the HART

®

standard menus. A manufac-

turer-specific DDD (Data-Device-Description)
is not necessary.

1) If you want to adjust a sensor with the PC, you have

to connect the PC cables to the signal cable outside
the Ex zone.
The intrinsic safety must not be impaired by this
connection. Use a suitable interface converter, e.g.
VEGACONNECT 2