VEGA VEGAWELL 72 4 … 20 mA User Manual

Page 32

Advertising
background image

D

eviation

-

T

urn down 1 : 1 up to 5 : 1

<

0.2 %

-

T

urn down up to 10 : 1

<

0.04 %

x TD

Influence of the product or ambient temperature
S

pecifications refer to the set span. Turn down (TD) = nominal measuring range/set

span.

Average temperature coefficient of the zero signal

I

n the compensated temperature range of 0 … +80 °C (+32 … +176 °F), reference

temperature 20 °C (68 °F).

A

verage temperature coefficient of the zero signal

-

T

urn down 1 : 1

<

0.15 %/10 K

-

T

urn down up to 5 : 1

<

0.2 %/10 K

-

T

urn down up to 10 : 1

<

0.25 %/10 K

O

utside the compensated temperature range:

A

verage temperature coefficient of the zero signal

-

T

urn down 1 : 1

typ. < 0.15 %/10 K

Thermal change, current output

T

hermal change, current output

<

0.15 %

at -40 … +80 °C (-40 … +176 °F)

Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)

S

pecifications refer to the set span. Turn down (TD) = nominal measuring range/set

span.

L

ong-term drift of the zero signal

< (0.1 %

x TD)/year

Total deviation (similar to DIN 16086)

T

he total deviation F

t

,

also called practical deviation, is the sum of the basic accuracy F

p

and long-term stability:

F

t

= F

p

+ F

s

F

perf

= √((F

T

)

2

+ (F

K

l

)

2

)

W

ith

-

F

t

: F

total

,

total deviation

-

F

p

: F

perf

,

basic accuracy

-

F

s

: F

stab

,

long-term drift

32

VEGAWELL

72 • 4 … 20

mA

9 S

upplement

27501

-EN

-080806

Advertising