VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART - two-wire User Manual

Page 57

Advertising
background image

1000

750

500

250

15

18

16

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 36: Voltage supply for version with increased sensitivity

1

HART load

2

Voltage limit EEx-ia instrument

3

Voltage limit non-Ex/Ex-d instrument

4

Operating voltage

Electrical protective measures

P

rotection, depending on housing version

-

P

lastic housing

IP

66/IP 67

-

A

luminium housing, stainless steel

housing - investment casting, stainless
steel housing - electro-polished

IP

66/IP 68 (0.2

bar)

12)

-

A

luminium and stainless housing, in-

vestment casting (optionally available)

IP

66/IP 68 (1

bar)

O

vervoltage category

III

P

rotection class

II

Functional safety (SIL)

F

unctional safety is already activated on instruments with SIL qualification ex factory. On

instruments without SIL qualification ex factory, the functional safety must be activated by the user
via the indicating and adjustment module or via PACTware for applications according to SIL.

F

unctional safety according to IEC 61508-4

13)

-

S

ingle channel architecture (1oo1D)

up to SIL2

-

double channel diversitary redundant
architecture (1oo2D)

up to SIL3

Y

ou can find detailed information in the supplied Safety Manual of VEGAPULS 63 or under "www.

vega.com", "Downloads", "Approvals".

12)

A

suitable cable is the prerequisite for maintaining the protection rating.

13)

N

ot with temperatures < -40 °C (-40 °F)

VEGAPULS

63 • 4 … 20

mA/HART - two-wire

57

10 A

nnex

28436

-EN

-120530

Advertising