Lock/unlock keys, Integration time and output characteris- tics – VEGA VEGADIF 34 … 51 User Manual

Page 37

Advertising
background image

VEGADIF 34 … 51

37

20094-EN-030731

Set-up

Display bias pressure

1) Push the z+ and s+ keys once simultane-

ously.

2) The adjusted bias pressure is displayed.
3) The display resets automatically to operat-

ing mode after approx. 2 s.

Save/accept bias pressure

1) Push the z+ and s+ keys twice simultane-

ously. The available pressure value is
accepted as bias value.

2) The display resets automatically to operat-

ing mode after approx. 2 s and displays
0.000 bar. The acting pressure (0 bar)
was thus accepted as bias pressure
(correction value).

Delete bias pressure

1) Push the z– and s– keys simultaneously

twice.

2) The display resets automatically to operat-

ing mode after approx. 2 s and displays
the value of the uncorrected bias pres-
sure.

Lock/unlock keys

Lock

1) Push the z+ and s– keys simultaneously

twice.

2) „Prot“ = protected appears.
3) The display resets automatically to operat-

ing mode after approx. 2 s.

Unlock

1) Push the z– and s+ keys simultaneously

once.

2) „Free“ = unlocked appears.
3) The display resets automatically to operat-

ing mode after approx. 2 s.

Integration time and output characteris-
tics

An integration time of 0, 1, 2, 4, 8, 16 s can be
adjusted via the step switch on your VEGADIF
to smooth the output signal. The output char-
acteristics for pressure, level and flow can be
selected as linear or extracted by root.

The integration time is the time required by the
output to reach 63 % of the actual height after
a signal jump of the differential pressure. After
three times the adjusted integration time, the
current output signal has reached 95 % of the
step change.

Explanation:
- an integration time of 0 s means tracking of

the current output without additional delay.

- an integration time of 16 s means that the

output will reach its final value only after
more than one minute.

Adjust the integration time and the output
characteristics acc. to the chart. The figures
indicate the switch positions:

Characteristics

linear

extracted by root

I ~

p

I ~

√∆

p/

p

max.

Integration time

0 s

1

9

0.5 s

2

A

1 s

3

B

2 s

4

C

4 s

5

D

8 s

6

E

16 s

7

F

Positions 0 and 8 allow the adjustment with the
PC (VVO) or the HART

®

handheld by which

you can also adjust the integration time of
0 … 40 s.

p

p

I

I

Advertising