VEGA VEGABAR 64 Foundation Fieldbus User Manual

Page 54

Advertising
background image

Output variable

Output

-

Signal

digital output signal, Foundation Fieldbus protocol

-

Physical layer

according to IEC 61158-2

Channel Numbers

-

Channel 1

Primary Value

-

Channel 2

Secondary Value 1

-

Channel 3

Secondary Value 2

-

Channel 4

Temperature Value

Transmission rate

31.25 Kbit/s

Current value

10 mA, ±0.5 mA

Dynamic behaviour output

Run-up time approx.

10 s

90 %

100 %

10 %

t

t

T

A

t

t

S

2

1

Fig. 28: Sudden change of the process variable, dead time t

T

, rise time t

A

and step response time t

S

1

Process variable

2

Output signal

Dead time

≤ 150 ms

Rise time

≤ 100 ms (10 … 90 %)

Step response time

≤ 250 ms (ti: 0 s, 10 … 90 %)

Damping (63 % of the input variable)

0 … 999 s, adjustable

Additional output parameter - temperature

Processing is made via output signal HART multidrop, Profibus PA and Foundation Fieldbus

Range

-50 … +150 °C (-58 … +302 °F)

Resolution

1 °C (1.8 °F)