Connection cable between IP 68 instrument and external housing:

-

Configuration

four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

-

Standard length

5 m (16.40 ft)

-

Max. length

180 m (590.5 ft)

-

Min. bending radius at 25 °C/77 °F

25 mm (0.985 in)

-

Diameter approx.

8 mm (0.315 in)

-

Colour - standard PE

Black

-

Colour - standard PUR

Blue

-

Colour - Ex-version

Blue

Indicating and adjustment module

Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection class

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Power supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-id instrument

9 … 32 V DC

Operating voltage with lighted indicating and adjustment module

-

Non-Ex instrument

12 … 32 V DC

-

EEx-ia instrument

12 … 24 V DC

-

EEx-id instrument

12 … 32 V DC

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

VEGABAR