3.2 Principle of operation

VEGABAR

64 is a pressure transmitter for use in the paper, food

processing and pharmaceutical industries as well as in water/sewage
water plants. Depending on the version, it is used for level, gauge,
absolute pressure or vacuum measurement. Measured products are
gases, vapours and liquids, also those containing abrasive substan-
ces.

The sensor element is the CERTEC

®

measuring cell with flush,

abrasion resistant ceramic diaphragm. The hydrostatic pressure of the
medium or the process pressure causes a capacitance change in the
measuring cell via the diaphragm. This change is converted into an
appropriate output signal and outputted as measured value.

The CERTEC

®

measuring cell is also equipped with a temperature

sensor. The temperature value can be displayed via the indicating and
adjustment module or processed via the signal output.

Power is supplied via the H1 Fieldbus. A two-wire cable according to
Fieldbus specification serves as carrier of both power and digital data
for multiple sensors. This cable can be operated in two versions:

l

via an H1 interface card in the control system and additional power
supply

l

via a Linking device
with HSE (High speed
Ethernet) and addition-
al power supply ac-
cording to IEC 61158-2

The DD (Device Descriptions) and CFF (capability files) necessary for
planning and configuration of your FF (Foundation Fieldbus) commu-
nication network are available in the download area of the VEGA
homepage www.vega.com under "Services - Downloads - Software -

Application area

Functional principle

Power supply and bus
communication

DD/CFF

VEGABAR

64 • Foundation Fieldbus

9