3 exchange of the electronics – VEGA VEGASWING 51 - contactless electronic switch User Manual

Page 22

Advertising
background image

Error

Cause

Removal

The signal lamp
flashes alternately
red and green

Shortcircuit or
overload

Check the electrical connection

Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

7

.3 Exchange of the electronics

To exchange the electronics in case of failure, it is not necessary to
dismount the instrument.

Dismounting the electronics module can destroy the housing seal.
Therefore only open the instrument if you want to insert a new
electronics module. The housing seal is supplied with the electronics
module.

You need an electronics module type SWE50T or SWE50C. If you
want to use an electronics module with a different signal output (e.g.
transistor output SWE50T), you can download the suitable operating
instructions manual from our homepage under Downloads. Take note
of the specifications in the respective operating instructions manual.

Proceed as follows to exchange the electronics module:

1

Separate VEGASWING 51 from operating voltage

2

Loosen fixing screw (1) of the valve plug (2) with a wrench

3

Remove valve plug (2) according to drawing

4

Remove lateral fixing screw (7) with a crosstip screwdriver

5

Pull electronics module (4) carefully out of the housing (8)

6

Remove the plug of the connection cable (6) from the socket on
the electronics module (4)

7

Set the 16-step rotating switch (5) of the new replacement
electronics module (4) to the value of the defective electronics
module

8

Insert the connection cable (6) in the socket of the new electronics
module (4)

9

Insert electronics module (4) into the housing (8). Make sure that
the lateral thread on the electronics module is above the hole on
the housing (8)

10 Push the electronics module (4) flush into the housing (8)
11 Screw in the lateral fixing screw (7) with a crosstip screwdriver
12 Plug the valve plug (2) to the instrument, make sure that the profile

seal (3) is placed correctly

13 Tighten fixing screw (1) with a wrench

VEGASWING

51 is again ready for operation.

Reaction after fault rec-
tification

22

VEGASWING

51 • - contactless electronic switch

7 Maintenance and fault rectification

30212

-EN

-111006

Advertising