3
Disconnect the connection cables according to the
operating instructions manual of the respective sensor
4
Loosen the two holding screws with a screwdriver (Torx
size T 10 or slot size 4)
2
1
Fig. 3: Loosen the screws
1
Electronics module
2
Screws (2 pcs.)
5
Pull the previous electronics out by holding the opening
levers.
Note:
Make sure that the housing is not rotated during the electronics
exchange. Otherwise the plug may be in a different position
later.
6
Insert the new oscillator carefully. Make sure that the plug
is in the correct position.
7
Insert the two hold screws and tighten them
8
Connect the connection cables according to the operating
instructions manual of the respective sensor
9
Screw the housing cover back on
The electronics exchange is now finished.
As a rule, the exchange of the electronics must be
documented internally when used in Ex applications.
10
Electronics module - VEGASON series 60
Mounting
30206
-EN
-070510