5 mounting procedure, four–wire instruments – VEGA VEGASON series 60 Electronics module User Manual
Page 11
4
.5 Mounting procedure, four-wire instruments
The electronics module is located in the electronics compart-
ment. The connection to the connection compartment is
provided via internal connection cables. The illustration shows
the position of the electronics compartment in the double-
chamber housing.
1
Fig. 4: Double chamber housing
1
Position of the electronics compartment
Proceed as follows:
1
Switch off power supply
2
Unscrew housing cover of the electronics compartment
3
Disconnect internal connection cables, on VEGASON 64,
65 and 66 also loosen the plug connector
4
Loosen the two holding screws with a screwdriver (Torx
size T 10 or slot size 4)
2
1
Fig. 5: Loosen the screws
1
Electronics module
2
Screws (2 pcs.)
5
Pull the previous electronics out by holding the opening
levers.
Mounting steps
Electronics module - VEGASON series 60
11
Mounting
30206
-
EN
-070510