5 instrument repair – VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual

Page 47

Advertising
background image

the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"S

ensor # online parameter adjustment". Connect this parameter

adjustment window before you carry out further steps.

Select with the right mouse key the sensor in the project and go to
"A

dditional function". Then click to "Software update". The window

"S

ensor # software update" opens. PACTware checks now the sensor

data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.

Push the button "Update software" and select the previously extracted
hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed
".

8

.5 Instrument repair

If a repair is necessary, please proceed as follows:

You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Downloads - Forms and certificates - Repair
form
".

By doing this you help us carry out the repair quickly and without
having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and, if need be, also a safety data
sheet outside on the packaging

l

Please ask the agency serving you for the address of your return
shipment. You can find the respective contact data on our website
www.vega.com under: "Company - VEGA worldwide"

Load software into sen-
sor

VEGAPULS

63 • Foundation Fieldbus

47

8 Maintenance and fault rectification

28451

-

EN

-120530