10 annex, 1 technical data, 10 .1 technical data – VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual

Page 49

Advertising
background image

10

Annex

10

.1 Technical data

General data

Materials, wetted parts

-

Process fitting

316L

-

Antenna encapsulated or hygienically
encapsulated

PTFE (TFM 1600), PFA plated, PVDF

-

Process seal with hygienically encapsu-
lated antenna

FKM (Viton), EPDM

Materials, non-wetted parts

-

Housing

Plastic PBT (polyester), Alu die-casting powder-
coated, 316L

-

Seal between housing and housing
cover

NBR (stainless steel housing), silicone (Alu/plastic
housing)

-

Inspection window in housing cover for
PLICSCOM

Polycarbonate (UL-746-C listed)

-

Ground terminal

316Ti/316L

Torque of the flange screws (min.)

60 Nm (44.25 lbf ft)

Weight

-

Clamp, bolting, depending on size and
housing material

3.5 … 6.0 kg (7.716 … 13.22 lbs)

-

Flanges, depending on flange size and
housing material

4.2 … 15.4 kg (9.259 … 33.95 lbs)

Output variable

Output

-

Signal

digital output signal, Foundation Fieldbus protocol

-

Physical layer

according to IEC 61158-2

Cycle time

min. 1 s (dependent on the parameter setting)

-

Damping (63 % of the input variable)

0 … 999 s, adjustable

-

Fulfilled NAMUR recommendations

NE

43

Channel Numbers

-

Channel 1

Primary value

-

Channel 2

Secondary value 1

-

Channel 3

Secondary value 2

Transmission rate

31.25 Kbit/s

Current value

10 mA, ±0.5 mA

Resolution, digital

> 1

mm (0.039 in)

VEGAPULS

63 • Foundation Fieldbus

49

10 Annex

28451

-

EN

-120530

Advertising