VEGA VEGAFLEX 67 (-200…+400°C) Profibus PA User Manual

Page 14

Advertising
background image

keep in mind that measurement down to the tip of the probe is not
possible. The exact value of the min. distance (lower dead band) is
stated in chapter "Technical data".

Fig. 4: Vessel with conical bottom

Make sure that the probe is not subjected to strong lateral forces.
Mount VEGAFLEX 67 at a position in the vessel where no
disturbances, e.g. from filling openings, agitators, etc., can occur.

Fig. 5: Lateral load

Extreme vibration caused by the system, e.g. due to agitators or
turbulence in the vessel from inflowing medium, can cause the coax
probe of VEGAFLEX 67 to vibrate in resonance. With coax probes of

Inflowing medium

14

VEGAFLEX

67 • Profibus PA

4 Mounting

32313

-EN

-120329

Advertising