3 exchanging the electronics module, 4 software update – VEGA VEGAFLEX 67 (-200…+400°C) Profibus PA User Manual

Page 39

Advertising
background image

Status accor-
ding to NE 107

Error code

Text message

Cause/Rectification

E

042

Hardware error, electronics defec-
tive

Exchange the instrument or send it
in for repair

E

043

Hardware error, electronics defec-
tive

Exchange the instrument or send it
in for repair

E

113

Communication conflict

Exchange the instrument or send it
in for repair

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.

In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.

If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.

The new electronics module must be loaded with the settings of the
sensor. These are the options:

l

At the factory by VEGA

l

Or on site by the user

In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module
").

The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.

Electronics module FX-E.67P suitable for VEGAFLEX 67 - Profibus
PA:

l

FX-E.

67PX (X = without approvals)

l

FX-E.

67PC (C = approvals XM, CX, CM according to product list)

8.4 Software update

The software version of VEGAFLEX 67 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

Sensor serial number

Assignment

Profibus PA

VEGAFLEX

67 • Profibus PA

39

8 Maintenance and fault rectification

32313

-

EN

-120329