12 Supplement

VEGAMET 391 • With SIL qualification

38704-EN-130228

Max. sampling rate

10 Hz

Relay outputs
Quantity

4 x operating relay

Function

Switching relay for level (relay 1 … 4)
Fault signal or pulse relay for flow/sampling pulse (relay

1/2)

Contact

Floating spdt

Contact material

AgSnO2, hard gold-plated

Turn-on voltage

min. 10 mV DC, max. 250 V AC/60 DC

Switching current

min. 10 µA DC, max. 3 A AC, 1 A DC

Breaking capacity

min. 50 mW, max. 500 VA, max. 54 W DC (with U less

than 40 V)

1)

Min. programmable switching hysteresis 0.1 %
Mode pulse output (relay 1/2)

Ʋ Pulse length

350 ms

Current output
Quantity

1 x output

Function

Current output for level

Range

4 … 20 mA

Resolution

1 µA

Max. load

500 Ω

Fault message

0 mA

Accuracy

Ʋ Standard

±20 µA (0.1 % of 20 mA)

Ʋ with EMC interferences

±200 µA (1 % of 20 mA)

Temperature error relating to 20 mA

0.005 %/K

USB interface

2)

Quantity

1 x on front plate

Plug connection

Mini-B (4-pole)

USB specification

2.0 (Fullspeed)

Max. cable length

5 m (196 in)

Indications
Measured value indication

Ʋ graphic-capable LC display

(65 x 32 mm), lighted

digital and quasianalogue indication

Ʋ Max. indicating range

-99999 … 99999

LED displays

1)

If inductive loads or stronger currents are switched through, the gold plating on the relay contact surface will be

permanently damaged. The contact is then no longer suitable for switching low-level signal circuits.

2)

Limited temperature range, see ambient conditions

Advertising