4 Mounting

4.1 General instructions

M

ake sure that all parts of the instrument in contact with the measured

product, especially the sensor element, process seal and process

tting, are suitable for the existing process conditions such as process

pressure, process temperature as well as the chemical properties of
the medium.

Y

ou can find the specifications in chapter "Technical data" in the or on

the type label.

S

elect an installation position you can easily reach for mounting and

connecting as well as later retrofitting of an indicating and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the indicating and adjustment module in four
different positions (each displaced by 90°).

U

se the recommended cables (see chapter "Connecting to power

supply") and tighten the cable gland.

Y

ou can give your instrument additional protection against moisture

penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This
applies mainly to outdoor mounting as well as installation in areas
where high humidity is expected (e.g. through cleaning processes) or
on cooled or heated vessels.

Fig. 2: Measures against moisture penetration

T

he ventilation of the measuring cell electronics is realised by a filter

element in the socket of the electronics housing. The ventilation of the
electronics housing is realised via an additional filter element around
the cable glands.

Suitability for process
conditions

Mounting position

Moisture

Ventilation

VEGABAR

52 • 4 … 20

mA/HART - climate compensated

11