VEGA VEGABAR 52 - climate compensated User Manual

Page 48

Advertising
background image

Nominal range

Overload, max. pressure Overload, min. pressure

-15 … 25

psig

700

psig

-15

psig

-15 … 70

psig

950

psig

-15

psig

-15 … 150

psig

1300

psig

-15

psig

-15 … 350

psig

1900

psig

-15

psig

-15 … 900

psig

2900

psig

-15

psig

-1.5 … 1.5

psig

500

psig

-15

psig

-3 … 3

psig

500

psig

-12

psig

-7 … 7

psig

700

psig

-15

psig

Reference conditions and actuating variables (similar to DIN EN 60770-1)

R

eference conditions according to DIN EN 61298-1

-

T

emperature

+

15 … +25 °C (+59 … +77 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa (12.5 … 15.4 psig)

D

etermination of characteristics

L

imit point adjustment according to IEC 61298-2

C

haracteristics curve

linear

R

eference installation position

upright, diaphragm points downward

I

nfluence of the installation position

<

0.2

mbar/20 Pa (0.003 psig)

Deviation determined according to the limit point method according to IEC 60770

8)

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA. Specifications

refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.

D

eviation

< (0.05 + 0.025

x TD)%

Influence of the product or ambient temperature

Average temperature coefficient of the zero signal

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA. Specifications

refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.

Nominal measuring range in
bar/kPa

Nominal
measuring
range in psig

Average temperature coeffi-
cient of the zero signal

9)

Average temperature coeffi-
cient of the zero signal

10)

0 … 0.2

bar/0 … 20 kPa

0 … 3

psig

< (0.05 + 0.05

x TD)%/10K

< (0.05 + 0.1

x TD)%/10K

0 … 0.4

bar/0 … 40 kPa

0 … 6

psig

< (0.05 + 0.05

x TD)%/10K

< (0.05 + 0.1

x TD)%/10K

0 … 1

bar/0 … 100 kPa

0 … 15

psig

< (0.02 + 0.05

x TD)%/10K

< (0.02 + 0.1

x TD)%/10K

0 … 2.5

bar/0 … 250 kPa

0 … 35

psig

< (0.02 + 0.05

x TD)%/10K

< (0.02 + 0.1

x TD)%/10K

0 … 5

bar/0 … 500 kPa

0 … 70

psig

< (0.02 + 0.03

x TD)%/10K

< (0.04 + 0.06

x TD)%/10K

48

VEGABAR

52 • 4 … 20

mA/HART - climate compensated

10 S

upplement

35075

-EN

-081209