VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Foundation Fieldbus Approval according to LPR radio standard User Manual
Page 49
Advertising
The following data or settings for adjustment of the indicating and
adjustment module are saved:
l
All data of the menu "Setup" and "Display"
l
In the menu "Additional settings" the items "Sensor-specific units,
temperature unit and linearization"
l
The values of the user programmable linearisation curve
The function can also be used to transfer settings from one instrument
to another instrument of the same type. If it is necessary to exchange a
sensor, the indicating and adjustment module is inserted into the
replacement instrument and the data are likewise written into the
sensor via the menu item "Copy sensor data".
VEGAPULS
61 • Approval according to LPR radio standard
49
6 Set up with the indicating and adjustment module