VEGA VEGABAR 81 Modbus and Levelmaster protocol - Operating Instructions User Manual

Page 57

Advertising
background image

57

10 Supplement

VEGABAR 81 • Modbus and Levelmaster protocol

46293-EN-131107

The specifications are only an overview and refer to the measuring cell. Limitations due to the

material and version of the process fitting as well as the selected pressure type are possible. The

specifications on the nameplate apply.

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +5 psig

+25 psig

-14.5 psig

0 … +15 psig

+75 psig

-14.5 psig

0 … +30 psig

+120 psig

-14.5 psig

0 … +150 psig

+450 psig

-14.5 psig

0 … +300 psig

+600 psig

-14.5 psig

0 … +600 psig

+1200 psig

-14.5 psig

0 … +1500 psig

+3000 psig

-14.5 psig

0 … +3000 psig

+6000 psig

-14.5 psig

0 … +9000 psig

+18000 psig

-14.5 psig

0 … +15000 psig

+22500 psig

-14.5 psig

-14.5 … 0 psig

+75 psig

-14.5 psig

-14.5 … +20 psig

+120 psig

-14.5 psig

-14.5 … +150 psig

+1200 psig

-14.5 psig

-14.5 … +300 psig

+1200 psig

-14.5 psig

-14.5 … +600 psig

+1200 psig

-14.5 psig

-3 … +3 psig

+25 psig

-14.5 psig

-7 … +7 psig

+75 psig

-14.5 psig

Absolute pressure
0 … +15 psi

+75 psi

0 psi

0 … +30 psi

+120 psi

0 psi

0 … +150 psi

+1200 psi

0 psi

0 … +300 psi

+1200 psi

0 psi

Output variable
Output

Ʋ Physical layer

Digital output signal according to standard EIA-485

Ʋ Bus specifications

Modbus Application Protocol V1.1b, Modbus over serial

line V1.02

Ʋ Data protocols

Modbus RTU, Modbus ASCII, Levelmaster

Max. transmission rate

19.2 Kbit/s

Dynamic behaviour output
Run-up time approx.

10 s