VEGA VEGABAR 81 Modbus and Levelmaster protocol - Operating Instructions User Manual

Page 58

Advertising
background image

58

10 Supplement

VEGABAR 81 • Modbus and Levelmaster protocol

46293-EN-131107

90 %

100 %

10 %

t

t

T

A

t

t

S

2

1

Fig. 32: Sudden change of the process variable. t

T

: dead time; t

A

: rise time; t

S

: jump response time

1 Process variable

2 Output signal

Dead time

≤ 45 ms

Rise time

≤ 35 ms (10 … 90 %)

Step response time

≤ 80 ms (ti: 0 s, 10 … 90 %)

Damping (63 % of the input variable)

0 … 999 s, adjustable

Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psi)

Determination of characteristics

Limit point adjustment according to IEC 61298-2

Characterstic curve

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

depending on the chemical seal version

Deviation (according to IEC 60770)
Turn down (TD) is the relation nominal measuring range/adjusted span.

Accuracy class

Non-linearity, hysteresis and re-

peatability with TD 1 : 1 up to 5 : 1

Non-linearity, hysteresis and re-

peatability with 5 : 1

0.2 %

< 0.2 %

< 0.04 % x TD

Influence of the product or ambient temperature
Thermal change zero signal and output span
Turn down (TD) is the relation nominal measuring range/adjusted span.

Average temperature coefficient,

zero signal

In the compensated tem-

perature range 10 … +70 °C

(+50 … +158 °F)

Outside the compensated tem-

perature range

Turn down 1 : 1

< 0.05 %/10 K

typ. < 0.05 %/10 K

Advertising