VEGA VEGABAR 81 Modbus and Levelmaster protocol - Operating Instructions User Manual

Page 60

Advertising
background image

60

10 Supplement

VEGABAR 81 • Modbus and Levelmaster protocol

46293-EN-131107

Isolating liquid

p

abs

> 1 bar/14.5 psi

p

abs

< 1 bar/14.5 psi

Version for oxygen ap-

plications

silicone oil KN2.2

-40 … +150 °C (-

40 … +302 °F)

-40 … +150 °C (-

40 … +302 °F)

-40 … +60 °C (-

40 … +140 °F)

Silicone oil KN2.2 with

cooling element

-40 … +200 °C (-

40 … +392 °F)

-40 … +150 °C (-

40 … +302 °F)

-40 … +60 °C (-

40 … +140 °F)

Silicone oil KN2.2 with

capillaries 2 m, 3 m

-40 … +200 °C (-

40 … +392 °F)

-40 … +150 °C (-

40 … +302 °F)

-40 … +60 °C (-

40 … +140 °F)

Silicone oil KN17

-90 … +180 °C (-

130 … +356 °F

-90 … +80 °C (-

130 … +176 °F)

-90 … +60 °C (-

130 … +140 °F)

High temperature oil KN32

with cooling element

-10 … +300 °C

(+14 … +572 °F)

-10 … +200 °C

(+14 … +392 °F)

-10 … +60 °C

(+14 … +140 °F)

High temperature oil

KN3.2 with cooling ele-

ment

-10 … +330 °C

(+14 … +626 °F)

+-10 … +200 °C

(+14 … +392 °F)

-10 … +60 °C

(+14 … +140 °F)

High temperature oil

KN3.2 with cooling ele-

ment 300 mm

-10 … +400 °C

(+14 … +752 °F)

+-10 … +200 °C

(+14 … +392 °F)

-10 … +60 °C

(+14 … +140 °F)

High temperature oil KN32

with capillaries 1 m, 2 m

or 5 m

-10 … +400 °C

(+14 … +752 °F)

+-10 … +200 °C

(+14 … +392 °F)

-10 … +60 °C

(+14 … +140 °F)

Halocarbon oil KN21

-40 … +150 °C (-

40 … +302 °F)

-40 … +80 °C (-

40 … +176 °F)

-

Halocarbon oil KN21 for

oxygen applications

-40 … +60 °C

(40 … +140 °F)

-40 … +60 °C

(40 … +140 °F)

-40 … +60 °C (-

40 … +140 °F)

Silicone-free liquid KN70

-40 … +70 °C (-

40 … +158 °F)

No vacuum

-40 … +60 °C (-

40 … +140 °F)

Med. white oil KN92 (FDA) -10 … +150 °C

(+14 … +302 °F)

-10 … +160 °C

(+14 … +322 °F)

-10 … +60 °C

(+14 … +140 °F)

Med. white oil KN92 (FDA)

with cooling element

-10 … +250 °C

(+14 … +482 °F)

-10 … +160 °C

(+14 … +322 °F)

-10 … +60 °C

(+14 … +140 °F)

Med. white oil KN92 (FDA)

with cooling element

300 mm

-

-10 … +160 °C

(+14 … +322 °F)

-10 … +60 °C

(+14 … +140 °F)

Mechanical stress, depending on the instrument version
Vibration resistance

1 to 4 g at 5 … 200 Hz according to EN 60068-2-6

(vibration with resonance)

Shock resistance

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Advertising